No concentrate draw
Cause
Solution

Concentrate container empty

 
Replace with full container

 


 

 

Excess foaming in discharge
Cause
Solution

Air leak in pick-up tube

 
Put clamp on tube or replace tube if brittle

 

 


 

 

Eclipse: Dead, no display
Controller connected to pump module via J1 cable?

Is power connected to pump module?

Is pump module circuit breaker tripped?

 


 

 

Eclipse: No pumps run

When pumps don't run it is important to differentiate between running on prime and running from a machine trigger. Pumps only run formula amounts from a machine trigger when pumps are calibrated.

When only certain pump positions are defective, check Pump Interface (PI) printed circuit board (PCB) and J1 cable connections.

Only one pump may run at a time in flush mode. Multiple pumps triggered at the same time will run in queue.

 


 

Eclipse: No pumps prime
Confirm pump motor harness is connected to PI PCB in pumpstand

Follow pumpstand manual troubleshooting procedures for PI PCB defect

 

 


 

Eclipse: No pumps on trigger, prime OK
Are pumps calibrated?

Is a delay time programmed or is pump in cue?

Is there a trigger signal? Confirm by viewing lights on Machine Interface and measuring signal with a volt meter
Follow pump module manual troubleshooting procedures for MI defect
Test run pump amounts as described in step 5.4.2 or Program Mode

 


 

 

Eclipse: Load count errors
Quite a number of perceived load count errors may occur in the normal day-to-day operations of a laundry. A load is counted (to the current formula number) each time the highest pump number with a nonzero amount runs. In Relay mode a load is counted each time pump 1 is run.

 


 

Eclipse: Load counts low
Check washer trigger signals

Check that machine operator may be short cycling the washer

 

 


 

Eclipse: Load counts high
Check washer trigger signal for count pump running more than once each load

 


 

Sprite: Dead, no display
Confirm main power connection voltage with a volt meter. Check fuse if optional fuse kit is installed. Substitute transformer followed by the printed circuit board to determine fault.

 


 


Sprite: No detergent feed

NOTE: Confirm that detergent type (screen 27) is set correctly

 

7.3.1 Probe Mode

Temporarily disconnect probe wire with detergent signal on to force detergent feed.

If detergent feeds, check for probe scaling and check detergent setpoint setting.

If detergent does not feed, substitute pump motor (or solenoid coil) followed by the printed circuit board to determine fault

7.3.2 Probe-less Mode

Confirm that unit is in a detergent feed condition as determined by rinse/initial charge signal inputs and initial charge/dose adjustments.

If detergent does not feed, substitute pump motor (or solenoid coil) followed by the printed circuit board to determine fault.

 

 


 

Sprite: Excess detergent consumption

7.4.1 Probe Mode

Check probe for scaling and measure wash solution via titration kit or by other means. If detergent us at proper concentration, troubleshoot the dish machine for clogged drains, excessive fresh water feedback, etc.

 

7.4.2 Probe-less Mode

Confirm that initial charge is dispensed only one time per initial fill. if you experience multiple initial charge feeds (Automatic/Conveyor mode only) review initial charge wiring connection. All other modes, adjust initial charge and dose adjustments accordingly.

 


 

Sprite: No rinse feed

Rinse feed is locked out under the following conditions:

Probe-less/Door/Automatic Initial Charge modes - Rinse is locked out after Initial Charge begins until the following rack (rinse signal) occurs.

Probe Mode - detergent signal did not occur within 90 seconds prior to rinse signal (Rinse Saver feature).

If not in one of the above conditions and rinse does not feed, substitute pump motor followed by the printed circuit board to determine fault.

 

 

 


 

UltraDose: Unit is not pumping fluid

Is the motor turning?

YES

  • check level of drain treatment supply.
  • check for air leaks in the supply tubing connection to pump tube stems.
  • check for clogs in inlet and outlet tubes.
  • insure pump tube is not at all twisted.
  • check pump tube. As it wears out, the amount of fluid pumped decreases. Change tube.

NO

  • ensure programmer is functioning: Try to run pump manually (see page 3 of the instruction manual).
  • check program: Be sure both switches are in center position for programmed dispensing.
  • check to insure wires from the electronic control to the motor are secure.
  • ensure that wire from batteries to electronic control is secure.
  • check battery condition. Try new batteries. BE SURE TIMER IS OFF before changing the batteries.
  • remove the pump to see if pump is binding the motor (run motor manually -- see page 3 of the instruction manual).

 

 


 

 

 

UltraDose: Unit is pumping too much fluid

Is the tubing the proper ID for your application?

NO

  • replace tubing with proper size.

YES

  • check program to insure the required duration.
  • ensure programmer is turning on and off at set times.
  • remember that fresh batteries will result in a slightly higher than average dispense.


 

 


 

UltraDose: Program time is locked up


Press reset. Reprogram timer.